AIM-M100 Medical Insulation Monitor Device
Technical Parameter
AUX Power | Voltage | AC220V±10% | Temperature monitoring | Thermal resistor | 2 Pt100 |
Power consumption | ≤8VA | Measuring range | -50~+200℃ |
Insulation monitoring | Resistance measuring range | 10~999kΩ | Alarm value range | 0~+200℃ |
Response value | 50~999kΩ | Alarm output | Output mode | 2 Relays |
Relative uncertainty | ±10% | Contact rating | AC 250V/3A DC 30V/3A |
Response time | ≤3s | Environment | Operating temperature | -10~+55℃ |
Permissible system leakage capacitance Ce | ≤5uF | Transport temperature | -25~+70℃ |
Measuring voltage Um | ≤12V | Storage temperature | -25~+70℃ |
Measuring current Im | ≤50uA | Relative humidity | 5%-95%,No condensation |
Impedance Zi | ≥200kΩ | Altitude | ≤2500m |
InternalDC resistance Ri | ≥240kΩ | IP degree | IP30 |
Permissible extraneous DC voltage Ufg | ≤DC280V | Rated impulse voltage / pollution degree | 4KV/Ⅲ |
Load current monitoring | Measuring Value | 2.1~50A | EMC | IEC 61326-2-4 |
Alarm Value | 5~50A | Communication | RS485(Modbus-RTU) |
Measuring accuracy | ±5% |