| Basic accuracy of CKT3544 series DC resistance tester is 0.02%, minimum resolution is 0.1μΩ, maximum test current is 1A, and the measurable range is 0.1μΩ~3.3MΩ. The test current is increased by the probe used to protect the port to achieve anti-interference measurement. Use the comparison judgment lamp (option) and the judgment sound of a large volume to accurately judge whether it is qualified or not at the production site. With EXT I/O interface suitable for NPN/PNP, it can be used in various automatic production lines. |
SPECIFICATIONS |
Model | CKT3544 | CKT3544-12 | CKT3544-24 |
No. of Channel | 1 | 12 | 24 |
Parameters | DC resistance |
Measurement Range | 0.1μΩ~3MΩ |
Min Resolution | 0.1μΩ |
Signal Source | 1A DC, Maximum DC 5.5 V |
Basic Accuracy | 0.1%±10 reading (3mΩ, 30mΩ, 3MΩ), 0.05%±4 reading (300kΩ), 0.02%±2 reading (other ranges) |
Range | 3mΩ/30mΩ/300mΩ/3Ω/30Ω/300Ω/3kΩ/30kΩ/300kΩ/3MΩ |
Speed | Fast (50Hz: 21ms, 60Hz: 18ms), slow (200ms) |
Temperature Compensation | Range: -10℃~60℃, Accuracy:±1℃ |
Display Max | 32, 000 |
Adjustment | Short-circuit clear zero for all ranges |
Comparator | 10 bins sorting, output HIGH/IN/LOW sorting signal |
Trigger | Internal trigger, External trigger, Manual trigger, BUS trigger |
Interface | External IO interface, LAN interface, RS232/485 interface |
Others | Temperature compensation function, comparator (ABS/REF%), lock button (OFF/menu lock/all lock), power frequency setting (AUTO/50Hz/60Hz), judgment tone setting, auto hold, average, panel save /read |
GENERAL FEATURES |
Power supply | AC100~256V, Frequency: 50/60 Hz, rated power: 15 VA |
Size | 325mm (L)x 215mm (W)x96.5mm (D); Weight: 3.5kg |
ORDER ING INFORMATION |
Accessories | KelvinTest leads; Temperature compensation probe; RS232 communication cable; AC power cord |