operated by the microcomputer system , which is able to inspect continuously
BAM 1006 Continuous Particulate Monitor consists of three basic components: the
instrument host, the cutting head, and sampling system. Each part is installed
individually for easy removal and replacement.
The functions of each part are described as follows:
1. Instrument host: The instrument host panel has display and buttons to realize
man-machine interact function. Internally integrated sampling system, mechanical
transmission control, signal detection and data processing, data transmission system.
2. Cutter: The cutter is designed according to the principle of aerodynamics and is
used for separation of particles of different diameters (PM10 and PM2.5), effective
flow of cutter cutting efficiency is 16.7L/min.
3. Sampling system: The sampling system mainly includes sampling pipeline,
dynamic heating system (DHS) and air pump. The air pump uses a wiper-type
atmospheric flow sampling vacuum pump, which has automatic lubrication, no oil
pollution, strong load capacity and other advantages.
Application
Mainly used in air quality monitoring networks, mobile monitoring stations,
environmental research, industrial and mining enterprises, research institutes and
other fields.
Features
● Beta-ray attenuation method to measure TSP, PM10, PM2.5 and PM1
●All the data is displayed on 7 inch LCD touch screen.
●Data storage: storage capacity can reach to 1, 000, 000 data. On board storage for
over 3 years of data
● Keyboard input: easy to input, man-machine dialogue mode
● One channel for sampling and analysis in case of any error caused by paper feed.
●Automatic fault diagnosis and alarms, also capable of fixing errors remotely
● Multi interface output: analog output, current output, relay output
●Very low operating cost
Advantages
Technology
Beta attenuation mass monitors are indifferent to the chemical composition of the particulate
matter whose mass density is being measured. As the name implies, beta attenuation mass
monitors measure only mass. No assumptions concerning the physical or optical properties of
the sampled particulate matter are necessary. This translates into reliable mass measurement
any time at any place.
Reliability
Over 5000 instruments are working in different environments from big cities to remote
villages .The BAM 1006 has a field-proven record of reliability even in the most challenging
environments and particulate concentrations.
Continuous Improvement
BAM 1006 has benefited from continuous improvement. Over time, functionality,
performance and reliability have advanced. The latest generation of BAM 1006
incorporates a new generation of microprocessor and advanced electronics allowing remote
operation, enhanced communication.
Touch Screen Display
Standard desk top capability allows operational checks, data validation and reporting from
distant locations. Using the supplied software control of the BAM 1006 is simple, which
presents a graphical representation of the actual BAM 1006 screen display with full remote
control. The front door mounted high-visibility touch screen maintains all of the existing
functionality of the previous generation of BAM 1006 in a modern, logical and intuitive
format for easy menu navigation.