-Monitors for both neonates and adults.
-Easy interfacing optional RS-232C (ICOM1) or USB (ICOM2) communication.
-Compact size 43x23x5.5 mm.
-Digital signal processing enhances performance under low perfusion,
patient motion and other signal interference.
Power Source; DC5Volt ± 5%(Option DC3.3V), 100mA
SpO2 Accuracy; ±1%(90-100% ), ±2%(80-89% ), ±3%(60-79% )
Interfacing; UART TTL or USB COM Port
Respons Time; ≤4sec, in case of Pulse Rate 80bpm
Protocol
-. UART [TTL Level (0~5[V])]
-. Asynchronous Operation
-. Baud rate : 38400bps
-. 8 Data bit
-. 1 Stop bit
-. no parity bit
Communication protocol
-. STX = 0xFA
-. ETX = 0xFB
-. Total 11byte Transmit
-. STX + Wave(2byte) + Hr(2byte) + Spo2(2byte) + Bar data(1byte)
+ Wave gain(1byte) + Status(1byte) + ETX
Communication data range
Wave=0 ~ 5000 (sensor or finger is open = 0x2500)
Hr=0 ~ 255 (sensor or finger is open = 0x9999)
SpO2=0 ~ 100( sensor or finger is open = 0x9999)
Bar data=0~10
Wave gain=0~10
Status= refer to the Status