Product Name | H8 Vital sign Patient Monitor |
ECG | Lead mode: 3-lead or 5-lead Lead selection: I, II, III, aVR, aVL, aVF, V Waveform: 5-lead, dual-channel 3-lead, single-channel Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s 2 RESPARATION Method Impedance between R-F(RA-LL) Differential Input Impedance: >2.5 MΩ Measuring Impedance Range:0.3~5.0Ω Base line Impedance Range:0 2.5 KΩ Bandwidth: 0.3 ~ 2.5 Hz |
NIBP | Method Oscillometric Mode Manual, Auto, STAT Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 (Min) Measuring Period in STAT Mode 5 Min Pulse Rate Range 40 ~ 240 bpm Alarm Typ: SYS, DIA, MEAN4.SpO2 Measurement and alarm range: 0 ~ 100% Resolution: 1% Measurement accuracy: 70%~100%: ±2%; 0%~69%: unspecified |
PR | Measurement and alarm range: 30 ~ 250 bpm Measurement accuracy: ±2 bpm or ±2%, whichever is greater |
TEMP | Channel: dual-channel Measurement and alarm range: 0 ~ 50℃ Resolution: 0.1 ℃ Accuracy: ±0.1 ℃I.6 7 IBP Label:ART, PA, CVP, RAP, LAP, ICP, P1, P2 Measuring and alarm range ART : 0 ~ 300 mmHg PA: -6 ~ 120 mmHg CVP/RAP/LAP/ICP : -10 ~ 40 mmHg P1/P2: -10 ~ 300 mmHg Press Sensor Sensitivity: 5 uV/V/mmHg Impedance: 300-3000Ω Resolution: 1 mmHg Accuracy: +-2% or +- 1mmHg, which great Actualization interval: about 1 Sec. 7.EtCO2 Method: Sidestream or Mainstream Measuring Range: 0~150mmHg Resolution: 0~69 mmHg, 0.1 mmHg 70~150 mmHg, 0.25 mmHg Accuracy: 0~40 mm Hg ±2 mm Hg 41~70 mm Hg ±5% 71~100 mm Hg ±8% 101~150 mm Hg ±10% AwRR Range: 2~150 rpm AwRR Accuracy: ±1BPM Apnea Alarm: Yes |